BEAMLINE X29

 

VIRTUAL TOUR

 

X29 BEAMLINE SEES THE LIGHT

General Information

SCHEDULE
BEAMLINE OPERATING PROCEDURES
RESEARCH PROGRAM: macromolecular crystallography
RING: X-RAY
STATUS: OPERATIONAL
TIME DEDICATED TO NSLS GENERAL USER PROGRAM: 25%

PARTICIPATING INSTITUTIONS:
Case Western Reserve University
Brookhaven Biology Department

LOCAL CONTACTS:
Howard Robinson: (631) 344-4789; hhr@bnl.gov
Wuxian Shi: (631) 344-2099; wushi@bnl.gov

SPOKESPERSON:
Bob Sweet: (631) 344-3401; sweet@bnl.gov
Mark Chance: (216) 368-4406; mrc16@case.edu

 

CAPABILITIES

Energy Range
(keV)

Crystal Type

Bandwidth @ E=10 keV
(eV)

Flux @ E=11.47 keV, I=300mA
(photons / sec)

Size of Focus
(mm)

6 - 15 keV

Si (111) 3 5 x 1011

through a 100 micron x 160 micron aperture

0.2mm H x 0.1mm V

OPTICAL AND BEAMLINE CONFIGURATION

Mini-gap Undulator : 12.5 magnetic period and 54 poles; tuning ranges for the 1st harmonic of 3.5 - 5.5 keV, for the 2nd harmonics on axis of 5-10 keV and for the 3rd harmonic of 11-15 keV.
Be Window #1 : 0.5mm thickness and 10.9 m from the source.
Monochromator : Double crystals monochromator with the 1st crystal cryogenically cooled to 100K and the 2nd crystal sagittally bent providing horizontal focusing; 11.9m (1st crystal) from the source.
Mirror : Striped configuration with 1/3 ULE substrate (uncoated), 1/3 palladium and 1/3 platinum coated; 12.75 m from the source.
Be Exit Window : 0.2mm thickness and 14.0 m from the source

 

EXPERIMENTAL STATION APPARATUS

ADSC (Area Detector System Corporation) Q315 detector.
Crystal Logic large detector goniostat with mini-kappa drive.
Motorized defining aperture.
Oxford Cryostream 700 Series Cryocooler.
Robotic crystal changer.

 

BEAMLINE AND EXPERIMENTAL STATION CONTROL SYSTEMS

Beamline Control System : X-ray optics adjusted by 12 V servo motors utilizing a CAMAC-based E500 stepper controller and windows XP operating system with Labview software.
Experimental Station Control System : CBASS linked to NSLS undulator gap system through EPICS, beamline optics control system (Labview ) and the Protein Crystallography Database (PXdb).

 

DATA PROCESSING AND BACKUP

Data Processing Computers : a number of Pentium Pro computers running with Linux operating system.
Data Processing Software : HKL2000 and Mosflm; other crystallographic software such as SHELX suite, CCP4; PHENIX and COOT also available.
Data Backup : Rimage 2000i DVD robot; external hard disk connecting to Linux computers.

 

Updated: 14 May 2007.